Journal article

Experimental Investigation of Refractive Index Measurement of Common Solvents and Aqueous Solutions in the Infrared Wavelengths

Abstract

We proposed and experimentally demonstrated a fast and simple technique to measure refractive index of liquids. A commercial Optical Time-Domain Reflectometer insert ignore into journalissuearticles values(OTDR); is implemented for interrogating sensor tips from a remote location. The system uses Fresnel reflected light from standard single mode fiber tips as sensing points without requiring any chemical coating, delicate fiber components, and/or sophisticated architecture at the sensor head. We reported new measurements of refractive indices of common solvents and solutions at both 1550 nm and 1625 nm. The measurement system precision was found to be of the order of 10 -4 . The experimental results have been validated by the way of theoretical calculations.

Keywords

Fiber optic refractometerFresnel reflectionOTDR

160 views · 12 downloads