- Caucasian Journal of Science
- Volume:6 Issue:1
- A STUDY ON CALCULATION OF FULL ENERGY PEAK EFFICIENCY OF NAI(TL) DETECTORS USING POINT SOURCE
A STUDY ON CALCULATION OF FULL ENERGY PEAK EFFICIENCY OF NAI(TL) DETECTORS USING POINT SOURCE
Authors : Gülçin BİLGİCİ CENGİZ EKER, İlyas ÇAĞLAR
Pages : 28-36
View : 19 | Download : 11
Publication Date : 2019-06-30
Article Type : Research Paper
Abstract :In this study, the energy dependence of efficiency and resolution of NaIinsert ignore into journalissuearticles values(Tl); detector were investigated using gamma ray spectrometry system. The full energy peak efficiency and energy resolution of NaIinsert ignore into journalissuearticles values(Tl); detector were experimentally measured using 152 Eu, 137 Cs, and 60 Co radioactive sources at the energy range of 204 keV to 1408 keV. It was found that that the efficiency and resolution of NaIinsert ignore into journalissuearticles values(Tl); detector decreases with the increasing gamma ray energy .Keywords : NaI Tl, gamma ray spectrometer, Full energy peak efficiency, Point source