Contrast-to-noise ratios of different dental restorative materials: An in-vitro cone beam computed tomography study

Abstract

Purpose In radiological views, strong beam hardening and streaking artifacts occur due to high-density structures and polyenergetic X-ray beams, and these lead to misdiagnosis. This study was performed in vitro to compare the contrast-to-noise ratio insert ignore into journalissuearticles values(CNR); of commonly used dental restorative materials by using Cone Beam Computed Tomography insert ignore into journalissuearticles values(CBCT); images with and without artifact reduction insert ignore into journalissuearticles values(AR); mode. Materials and Methods A total of 108 molar teeth were restored with nine different groups of restorative materials, with each group containing 12 teeth. Teeth were placed in a dry human mandible and scanned, one by one, via Planmeca 3D ProMax insert ignore into journalissuearticles values(Planmeca, Helsinki, Finland); with and without AR mode. Images were analyzed using ImageJ software insert ignore into journalissuearticles values(National Institutes of Health, Bethesda, MD); to calculate the CNR. Results CNR was calculated to be the highest in compomer insert ignore into journalissuearticles values(Glassiosite); images without AR mode insert ignore into journalissuearticles values(mean: 3.36); and with AR mode insert ignore into journalissuearticles values(mean: 3.61);. CNR was calculated to be the lowest in amalgam insert ignore into journalissuearticles values(Tytin); images without AR mode insert ignore into journalissuearticles values(mean: 0.21); and with AR mode insert ignore into journalissuearticles values(mean: 0.23);. A significant difference was found between materials in terms of CNR measurements insert ignore into journalissuearticles values(p ≤ 0.05);. CNR measurements were increased after the AR mode application insert ignore into journalissuearticles values(p ≤ 0.05);. Conclusion AR mode was effective in reducing artifacts arising from dental materials on CBCT images, so it is necessary to use AR mode for correct diagnoses.

Keywords

Cone Beam Computed Tomographyontrast to noise ratiodental materialsartifactsimage quality

40 views · 12 downloads