Journal article

Evaluation of the Accuracy of Output Values of Torque Wrenches in Clinical Settings

Abstract

Aim: The aim of this in vitro study was to determine and compare the accuracy of different types of mechanical torque limiting devices currently used in clinical settings. Materials and Methods: A total of 30 torque wrenches including both friction-style (n=17) and spring-style (n=13) instruments were evaluated in this study. A calibrated digital torque meter (Lutron TQ-8800; Lutron Electronic Enterprise Co., Ltd.; Taiwan) was used to measure the output torque values of the torque wrenches. Measurements were repeated 10 times for each wrench and reported in Ncm. The Shapiro-Wilk test was performed for mean torque measurment values followed by an independent t-test and the percentages of absolute deviation from target torque values were calculated. The results were evaluated at a 95% confidence interval with significance at p 0.05). While the percentage deviation values ranged between 7.64 and 11.76 for friction-style devices, they were 1.79- 5.12 for spring-style devices. Conclusion: The accuracy of spring-type torque wrenches is better than that of friction-type torque wrenches, and the calibration of all wrenches should routinely be performed at the target torque value specified by the manufacturer to ensure optimum healthcare services.

Keywords

Torklu raşetDental implantMekanik tork sınırlayıcı aletKomplikasyonKırık