- Determination of the Radiation Dose Level in Different Slice Computerized Tomography
Determination of the Radiation Dose Level in Different Slice Computerized Tomography
Authors : Osman GÜNAY, Özcan GÜNDOĞDU, Mustafa DEMİR, Mohammad ABUQBEİTAH, Doğan YAŞAR, Serpil AKÖZCAN, Enis KAPDAN, Onur YARAR
Pages : 119-123
Doi:10.22399/ijcesen.595645
View : 24 | Download : 15
Publication Date : 2019-11-30
Article Type : Research Paper
Abstract :Today, many imaging tools have emerged in the medical field using ionizing radiation. Computed tomography insert ignore into journalissuearticles values(CT); is regarded the most creative and common modality among the other imaging devices. In CT imaging, radiation may scatter to the environment from the patient and the interaction medium. In this study, two CT scanners with different slice numbers were used to measure the scatter radiation dose to the environment. Alderson rando phantom was utilized in the acquired CT exams as human body simulation. Radiation dose levels at different distances were determined by placing thermoluminescence dosimeters in the surrounding environment insert ignore into journalissuearticles values(at different distances); during phantom’s head-neck and thorax CT imaging. As a result, it has been found that the radiation dose varies between 13.90±1.96 µSv and 96.79±11.12 µSv in head and neck CT imaging. While, in thorax CT scan, the radiation dose varies between 5.28±0.84 µSv and 20.63±1.76 µSv.Keywords : CT, Radiation Dose, TLD, X ray