Performance Comparison between Monolithic C18 and Conventional C18 Particle-Packed Columns in the Liquid Chromatographic Determination of Propranolol HCl

Authors : Sami El DEEB, Hermann WÄTZIG
Pages : 543-552
View : 19 | Download : 7
Publication Date : 0000-00-00
Article Type : Research Paper
Abstract :Monolithic and conventional particle-packed columns were applied for the determination of propranolol hydrochloride in the presence of its 2 main degradation products, 3-insert ignore into journalissuearticles values(1-naphthyloxy);-propane-1,2- diol and 4-isopropyl-1,7-bis-insert ignore into journalissuearticles values(1-naphthyloxy);-4-azaheptane-2,6-diol. The separations were investigated on monolithic columns at flow rates from 1 to 9 mL/min. Fast and efficient separation was obtained by monolithic columns. The analysis time was decreased by about 5-fold on monolithic columns at a flow rate of 4 mL/min, while maintaining sufficient resolution between propranolol and its degradation products. The method was validated using a set of 3 monolithic columns and compared to a conventional insert ignore into journalissuearticles values(Superspher); C18 column. The precision for both retention time and peak area was investigated over a wide concentration range insert ignore into journalissuearticles values(0.002- 1 mg/mL); and found to be equal or slightly better on Chromolith Performance compared to the conventional column. Batch to batch reproducibility of the Chromolith Performance columns insert ignore into journalissuearticles values(n = 3); was also calculated. The RSDs % equal 0.66% for retention time and ranged from 0.45% to 1.12% for peak areas. Practical parameters including the pressure drop, plate height, retention time and resolution of monolithic columns were compared to those of a conventional insert ignore into journalissuearticles values(Superspher); C18 column. The detection and quantitation limits on monolithic columns at both flow rates insert ignore into journalissuearticles values(1 and 4 mL/min); were 0.012 and 0.04 m g/mL, compared with 0.061 and 0.2 m g/mL on the conventional column. The method showed good linearity and recovery and was found to be suitable for the analysis of propranolol hydrochloride formulations.
Keywords : Monolithic columns, Chromolith Performance, propranolol hydrochloride, performance, validation, degradation products, method transfer, batch reproducibility

ORIGINAL ARTICLE URL
VIEW PAPER (PDF)

* There may have been changes in the journal, article,conference, book, preprint etc. informations. Therefore, it would be appropriate to follow the information on the official page of the source. The information here is shared for informational purposes. IAD is not responsible for incorrect or missing information.