- Turkish Journal of Electrical Engineering and Computer Science
- Volume:25 Issue:1
- Analysis of reconstruction performance of magnetic resonance conductivity tensor imaging (MRCTI) usi...
Analysis of reconstruction performance of magnetic resonance conductivity tensor imaging (MRCTI) using simulated measurements
Authors : EVREN DEĞİRMENCİ, BEHÇET MURAT EYÜBOĞLU
Pages : 474-483
View : 16 | Download : 8
Publication Date : 0000-00-00
Article Type : Research Paper
Abstract :Magnetic resonance conductivity tensor imaging insert ignore into journalissuearticles values(MRCTI); was proposed recently to produce electrical conductivity images of anisotropic tissues. Similar to magnetic resonance electrical impedance tomography insert ignore into journalissuearticles values(MREIT);, MRCTI uses magnetic field and boundary potential measurements obtained utilizing magnetic resonance imaging techniques. MRCTI reconstructs tensor images of anisotropic conductivity whereas MREIT reconstructs isotropic conductivity images. In this study, spatial resolution and linearity of five recently proposed MRCTI algorithms are evaluated using simulated measurements gathered from three different computer models. The results show that all five algorithms have quite similar reconstruction performances. Since the AB$_{z}$S algorithm is easier to apply compared to the other four algorithms it can be said to be the best algorithm among the five algorithms.Keywords : Anisotropic conductivity, electrical impedance, imaging, magnetic resonance, reconstruction